[2] In 1977, Jim Peacock of the Australian Academy of Science asked Bill Snowdon, then Chief CSIRO AAHL if he could have the newly released USA NIH and the British equivalent requirements for the development of infrastructure for bio-containment reviewed by AAHL personnel with a view to recommending the adoption of one of them by Australian authorities.
The Australian Centre for Disease Preparedness was commissioned in 1985 as the Australian Animal Health Laboratory,[3] includes a number of laboratories and animal house facilities capable of operating at three steps of negative pressure below the lowest pressure within any part of the extramural wind-zone of the building exterior.
External pressure and the cascaded negative pressure zones are measured and maintained in real time by electronic sensors, electronically controlled supply and exhaust air handling systems and purpose developed predictive logic controllers in support of management protocols to manage air locks, access and services to prevent loss or ingress of potentially contaminated gases, solids or liquids .
Supply air is protected by two 0.001 HEPA filters in tandem, (series mounted) with appropriate pre-filters in purpose designed individual housings.
Waste liquids can be dosed chemically in steam jacketed stainless steel tanks as the first barrier, followed by heating under pressure and maintained at elevated temperature for an appropriate time.
The Prime Minister Malcolm Fraser, although in favor of the project, threatened serious consequences for management if the design team "got it wrong".