Low temperatures may be maintained within a cryostat by using various refrigeration methods, most commonly using cryogenic fluid bath such as liquid helium.
Closed-cycle cryostats consume a relatively large amount of electrical power, but need not be refilled with helium and can run continuously for an indefinite period.
The liquid helium may be replenished as it boils away, at intervals between a few hours and several months, depending on the volume and construction of the cryostat.
In the older designs there may be additional liquid nitrogen bath, or several concentric layers of shielding, with gradually increasing temperatures.
In order to achieve temperatures lower than liquid helium at atmospheric pressure, additional cooler stages may be added to the cryostat.
Cryostats used in MRI machines are designed to hold a cryogen, typically helium, in a liquid state with minimal evaporation (boil-off).
If, for any reason, the wire becomes resistive, i.e. loses superconductivity, a condition known as a "quench", the liquid helium evaporates, instantly raising pressure within the vessel.
Nowadays few units using multiple cryogens are made with the trend being towards 'cryogen-free' cryostats in which all heat loads are removed by cryocoolers.
To minimize unnecessary warming all necessary mechanical movements of the microtome can be achieved by hand via a wheel mounted outside the chamber.
Once the specimen is cut to a satisfactory quality, it is mounted on a warm (room temperature) clear glass slide, where it will instantaneously melt and adhere.
The entire process from mounting to reading the slide takes from 10 to 20 minutes, allowing rapid diagnosis in the operating room, for the surgical excision of cancer.